WEKO3
アイテム
三軸加速度計を用いた振戦解析・評価システムの構築
http://hdl.handle.net/10649/606
http://hdl.handle.net/10649/606bc3d6504-17ea-42f1-89be-9556e654588d
| 名前 / ファイル | ライセンス | アクション |
|---|---|---|
|
|
|
| アイテムタイプ | 紀要論文 / Departmental Bulletin Paper(1) | |||||
|---|---|---|---|---|---|---|
| 公開日 | 2011-02-15 | |||||
| タイトル | ||||||
| タイトル | 三軸加速度計を用いた振戦解析・評価システムの構築 | |||||
| 言語 | ||||||
| 言語 | jpn | |||||
| キーワード | ||||||
| 主題Scheme | Other | |||||
| 主題 | Parkmson's disease | |||||
| キーワード | ||||||
| 主題Scheme | Other | |||||
| 主題 | tremor analysis | |||||
| キーワード | ||||||
| 主題Scheme | Other | |||||
| 主題 | 3-axis accelerometer | |||||
| 資源タイプ | ||||||
| 資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
| 資源タイプ | departmental bulletin paper | |||||
| 著者 |
松本, 義伸
× 松本, 義伸× 高井, 俊輔× 塩田, 宏嗣× 田村, 正人× 福本, 一朗 |
|||||
| 著者別名 | ||||||
| 識別子Scheme | WEKO | |||||
| 識別子 | 3479 | |||||
| 姓名 | MATSUMOTO, Yoshinoobu | |||||
| 著者別名 | ||||||
| 識別子Scheme | WEKO | |||||
| 識別子 | 3480 | |||||
| 姓名 | TAKAI, Syunsuke | |||||
| 著者別名 | ||||||
| 識別子Scheme | WEKO | |||||
| 識別子 | 3481 | |||||
| 姓名 | SHIOTA, Hiroshi | |||||
| 著者別名 | ||||||
| 識別子Scheme | WEKO | |||||
| 識別子 | 3482 | |||||
| 姓名 | TAMURA, Masato | |||||
| 著者別名 | ||||||
| 識別子Scheme | WEKO | |||||
| 識別子 | 3483 | |||||
| 姓名 | FUKUMOTO, Ichiro | |||||
| 著者所属 | ||||||
| 値 | 長岡技術科学大学ラジオアイソトープセンター | |||||
| 著者所属 | ||||||
| 値 | 長岡技術科学大学生物系 | |||||
| 著者所属 | ||||||
| 値 | 長岡西病院神経内科 | |||||
| 著者所属 | ||||||
| 値 | 日本大学医学部神経内科教室 | |||||
| 著者所属 | ||||||
| 値 | 長岡技術科学大学生物系 | |||||
| 抄録 | ||||||
| 内容記述タイプ | Abstract | |||||
| 内容記述 | We construct the tremor analysis and evaluation system using 3-axis accelerometers for diagnosis of Parkmson's disease This system consists of 3-axis accelerometers (TEAC Instruments Co , 612-ZS), an analog-digital signal conversion card (National Instruments Co , DAQ-Card6024E), a personal computer (SHARP Co , Mebius PC-CBI-Ml) and an analysis program We created the program by the measurement software (National Instruments Co , Lab View 6j) The upper limb tremor accelerations were measured to 25 Parkmson's disease patients, 9 essential tremor patients, and 12 healthy students using this system, which we built A sampling rate is 20msec and a measurement period is 30 seconds While we were recording the tremor, the subject was sitting on the chair and putting his elbows on the desk We calculated a tremor intensity and peak frequencies on the recorded upper limb tremor of the subject The tremor intensity was calculated from X, Y and Z signals of the 3-axis accelerometers The tremor peak frequency was calculated from the strongest axis signal The calculated parameters were compared with a subject's inside information, which are Age, Tremor history and Yahr's grade The result of this investigation, we found a tendency, which the right-and-left difference of the tremor intensity decreases as the Yahr's grade becomes bad This result suggests that this system is useful to the degree evaluation of serious illness of a Parkmson's disease patient or the diagnosis of a tremor disease patient | |||||
| 書誌情報 |
長岡技術科学大学研究報告 巻 24, p. 99-106, 発行日 2002-01-15 |
|||||
| ISSN | ||||||
| 収録物識別子タイプ | ISSN | |||||
| 収録物識別子 | 0388-5631 | |||||
| 書誌レコードID | ||||||
| 収録物識別子タイプ | NCID | |||||
| 収録物識別子 | AN00177120 | |||||
| 著者版フラグ | ||||||
| 出版タイプ | VoR | |||||
| 出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 | |||||
| その他のタイトル | ||||||
| その他のタイトル | Study on the construction of the tremor analysis and evaluation system using 3-axis accelerometers | |||||
| 出版者 | ||||||
| 出版者 | 長岡技術科学大学 | |||||